1. Daily inspection
Lamp working status (PRX6000 and 8000). Method: Use a piece of white paper on the work area, if there is an arc-shaped shadow, change the corresponding bulb. -The status of the disinfection handle in place. Method: Two "clicks" during installation. -Cleaning: Wash the exterior with a weak alkaline solvent (soap and water); avoid using chlorine-containing lotion (damages metal) and alcohol lotion (damages plastics and paint).
2. Monthly inspection
It is mainly to check whether the backup power system (battery) is normal. Method: Cut off the 220V power supply and see if the backup power supply is activated.
3. Lamp life
The average life span is 1000 hours; for the lamp holder, it is basically replaced once a year.
4. Annual inspection
Four maintenance steps (performed by a certified engineer)-tighten the power cord connector (input and output of the control box). Screws and brushes at each connection-Adjust the rotation limit (for example, 4000 type lamp cap surrounds its lamp cap arm) Lamp working voltage (23V AC+DC) Each joint brake-Check the verticality of the suspension pipe and the balance of the suspension system. Each part of the connection Whether the tightening of the screws is normal. Whether the brakes rotate normally when the joints are in motion. Whether the limit heat dissipation effect. The status of the lamp holder. The status of the disinfection handle.